Basic Operation of FPP 5000
The VEECO FPP-5000 4-point probe simplifies the measurement
of resistive properties of semiconductor wafers and resistive films. The
microprocessor based electronics permits direct computation of V/I, sheet or
slice resistivity, and metallization thickness and P-N type testing. Unlike most
four point probes and probing stations, which move the probe head into the
wafer, the FPP-5000 is designed so that the wafer is moved into the probe head.
This insures constant probe force independent of operator force and wafer
thickness.
Keyboard Description
-
V/I - Displays
measured V/I multiplied by programmed GEOM Constant.
-
SHEET - Displays
sheet resistance based on measured V/I.
-
SLICE - Calculates
the bulk resistivity of a slice of layer whose thickness is entered in the
PRGM mode.
-
THICK - Calculates
the thickness of a layer of slices whose bulk resistivity is entered in the
PRGM mode.
-
TYPE - When
selected, a type test is performed.
-
PEN - When
selected, a penetrate voltage is applied to the sample.
-
PRGM - Put the
keyboard into the PRGM mode for entering bulk resistivity or film thickness.
Press STORE afterwards to store the input into system memory.
-
SELF-TEST -
Performs a system self test.
-
RETEST - Performs a
retest when the previous test failed.
-
CONST - Displays
the selected input stored in the memory.
-
Numerical Keys - To
enter the input required for calculation.
-
Wafer Measurements in General
1. Determine thickness of wafer
with micrometer (remember the thickness).
-
Adjust the rear stop by loosening the thumb screw and moving the
position of the stop to the appropriate wafer size position. (100mm,
125mm, 150mm, 2 inch, and 3 inch positions are in accordance with SEMI
STD MI. 1.)
-
Place the wafer you wish to probe in the appropriately sized wafer
holder.
-
Place the appropriate size backing plate with the spiral side facing the
wafer.
-
Slide the wafer holder forward so that it is "banking" on the front
stop. The wafer is now positioned to be probed in the center.
-
Ensure the power to the FPP 5000 is on. The power switch is located on
the left rear of the FPP 5000.
-
Select the SLICE mode and ensure the indicator light illuminates.
-
Depress the PRGM mode and ensure the indicator light illuminates.
-
Enter the wafer thickness in microns and depress the STORE key. The
display should wink.
-
Depress the PRGM mode and ensure the indicator light goes out.
-
Select the proper display mode. Each switch illuminates when you depress
the switch for the mode that you want. Depressing the switch a second
time causes the mode to be deselected. The modes are
V/I switch - Displays measured
V/I multiplied by programmed GEOM constant.
SHEET resistance switch -
Displays the resistivity based on the measured V/I.
SLICE resistivity displays
switch - Calculates the bulk resistivity of a slice or a layer whose thickness
is entered by means of the PRGM mode keyboard switches.
THICK switch - Calculates the
thickness of a layer or slices whose slice resistivity is entered on the PRGM
mode keyboard switches.
-
Close the cover and hold it down to initiate a test. The display will go
blank and the TEST indicator adjacent to the display will illuminate for
not more than 2 seconds at which time the display will reappear. E 02
will appear on the display if the cover is released before a measurement
is completed.
By depressing the RETEST button,
with the cover down, you can reinitiate the test. The display will indicate E 01
if the cover is not held down while pressing the retest button.
It is recommended that all
measurements be performed with the cover closed to eliminate effects due to
light and RFI
-
Open the cover.
-
Slide the wafer holder so that it "banks" against the rear stop and
rotate it so that the location marked "2" is facing you (90o).
-
Probe this location as was done in step 5.
-
Repeat steps 6 through 8 for locations marked 3, 4, and 5.
-
Remove the wafer backing plate and remove the wafer.
-
Determining a Layer Thickness
-
Press the THICK switch and observe that the THICK indicator illuminates.
-
Press the PRGM switch to put the instrument in the program mode. Observe
that the PRGM indicator and units indicator W - cm are illuminated.
-
Enter the slice resistance of the aluminum layer.
-
Press STORE and observe that the display winks.
-
Press the PRGM switch to exit from the program mode. Observe that the
PRGM indicator extinguishes.
-
Follow procedure A using the THICK mode.
-
Determine Layer Resistivity
-
Press the SLICE switch and observe that the SLICE indicator illuminates.
-
Press the PRGM switch to put the instrument in the program mode. Observe
that the PRGM indicator and the units indicator MIL are illuminated.
-
Enter the layer thickness in microns using the unit selection and data
entry switches.
-
Press the m key to change units from MILS to microns. Observe that the m
indicator illuminates.
-
Press STORE and observe that the display winks.
-
Press the PRGM switch to exit from the program mode. Observe that the
PRGM mode indicator extinguishes.
-
Follow procedure A ensuring that you use the SLICE mode.
-
Determining Layer Conductivity Type
-
Adjust the rear stop by loosening the thumb screw and moving the
position of the stop to the appropriate wafer size position. (100mm,
125mm, 150mm, 2 inch, and 3 inch positions are in accordance with SEMI
STD MI. 1.)
-
Place the wafer you wish to probe in the appropriately sized wafer
holder.
-
Place the appropriate size backing plate with the spiral side facing the
wafer.
-
Slide the wafer holder forward so that it is "banking" on the front
stop. The wafer is now positioned to be probed in the center.
-
Ensure the power to the FPP 5000 is on. The power switch is located on
the left rear of the FPP 5000.
-
Select the V/I switch and ensure the V/I indicator illuminates.
-
Close the cover and hold it down to initiate a test. The display will go
blank and the TEST indicator adjacent to the display will illuminate for
not more than 2 seconds at which time the display will reappear. E 02
will appear on the display if the cover is released before a measurement
is completed.
By depressing the RETEST button,
with the cover down, you can reinitiate the test. The display will indicate E 01
if the cover is not held down while pressing the retest button.
It is recommended that all
measurements be performed with the cover closed to eliminate effects due to
light and RFI
-
Press the TYPE switch and notice that the TYPE indicator
illuminates. The type of the wafer will be indicated on the display.
-
Open the cover.
-
Remove the wafer backing plate and remove the wafer.
-
Using the Pen Feature
-
The PEN function applies a High Voltage low energy pulse to the probe
tips in order to break down any thin oxide that has grown on the sample.
- To
enable the PEN function depress the PEN switch before step 7 in part
A. Ensure that the PEN indicator illuminates.
- To
disable the Pen function depress the PEN switch and ensure the PEN
indicator goes out.
|